   PRODUCT: 
SITRANS TF2, digital display thermometer



Code: 7NG3140
Product name: SITRANS TF2, digital display thermometer

It is used to indicate and monitor the temperature measured at the point of installation.

 

Measuring principle

 

Resistance thermometer

Pt100 class B acc. to DIN IEC 751

Input

 

Measured variable

Temperature

Max. measuring range

-50 °C ... +200 °C (-58 ... +392 °F)

Min. measured span

50 K (90 °F)

Output

 

Output signal

4 ... 20 mA, 2-wire

Lower current limit

Min. 3.6 mA

Upper current limit

Max. 23 mA

Output protected against

Reversed polarity, overvoltage and short-circuiting

Max. load

(UH – 12 V) / 0.023 A

Characteristic

Temperature-linear

Measuring accuracy

 

Error in measurement at 23 °C ± 5 K (73.4 ± 9 °F)

< ± (0.45 K + 0.2 % of full-scale value in K + 1 digit in K)
(< ± (0.81 °F + 0.2 % of full-scale valuein °F + 1 digit in F)

Measuring cycle time

≤ 100 ms

Temperature effect

< ± 0.15 %/10 K (< ± 0.15 %/18 °F)

Power supply effect

< ±0,01 % of full-scale value/V

Vibration influence

< ±0.05 %/g to 500 Hz in all directions (to IEC 68‑2‑64)

Rated operating conditions

 

Ambient conditions

 

Ambient temperature

-25 ... +85 °C (-13 ... +185 °F)

Temperature range for best readability

-10 ... +70 °C (14 ... 158 °F)

Storage temperature

-40 ... +85 °C (-40 ... +185 °F)

Degree of protection

IP65 to EN 60529

Electromagnetic compatibility

EN 61326/A2 Appendix A (2001)

Displays and controls

 

Display

LCD, max. 5 digits, digit height 9 mm (0.354 inch)

Resolution at max. measuring range

0.01 °C (0.01 °F)

Decimal point

Freely parameterizable

Limit values

Freely parameterizable

Limit violation display

Red LED and message on LCD
(↑ symbol /↓ symbol in case of limit violation in upward / downward direction)

Parameterization

With 3 keys

Units

mA or % or Ω or physical variable: °C, °F, °R, K

Damping

Between 0.1 and 100 s

(increment: 0.1 s) freely parameterizable

Design

 

Weight

≈ 0.7 kg (≈ 1.54 lb)

Non-wetted parts materials

 
  • Field housing

Diam. 80 mm (diam. 3.15 inch), stainless steel, mat. No. 1.4016

  • Cover

Stainless steel, mat. No. 1.4016 with glass

Wetted parts materials

 
  • Protective tube

To DIN 43772 form 8 (March 2000), Ø 14 x 1.5 mm (Ø 0.55 x 0.06 inch)

  • Material

Stainless steel (mat. No. 1.4571/316Ti)

  • Protective tube screw socket

G½B to DIN 3852-2 form A or ½“-14 NPT

  • Material

Stainless steel (mat. No. 1.4571/316Ti)

Measuring insert

Length to fit the ordered protective tube, stainless steel

Connection of display to the protective tube

Radial (type A), can be swiveled by max. ±120° (α)

 

axial (type B), can be swiveled by max. ±360°

Length of the protective tube(U1)

See Ordering data

Electrical connection

Using 2-pole plug connector made of plastic with M16x1.5‑cable entry to EN 175301‑803A, ½"-14 NPT

Power supply

 

Terminal voltage on temperature transmitter (UH)

12 ... 30 V DC

Operating limits

 

Pressure

Max. 40 bar (580 psi)

 

Share |
Video clips
Hot News
Certificates
Online Support
Sales Support
0901 185 205
Support via Skype     Support via Yahoo Messanger    
Technical Support
0938 77 40 99
Support via Skype     Support via Yahoo Messanger    
 
An Thinh Automation Ltd.,
26/50 Nguyen Minh Hoang street, Tan Binh District, HCM City, Viet Nam
Tel:(84) 08 3948 1715         Fax: (84) 08 3811 1332