   PRODUCT: 
SITRANS P200/P210/P220



Code: 7MF1565/7MF1566/7MF1567
Product name: SITRANS P200/P210/P220

The SITRANS P200 pressure transmitter measures the gauge and absolute pressure of liquids, gases and vapors.

  

Pressure transmitters for food, pharmaceuticals and biotechnology

Mode of operation

 

Measuring principle

Piezo-resistive

Input

 

Measured variable

Gage or absolute pressure

Measuring range

0 ... 160 mbar (0 ... 2.32 psi)
...
0 ... 40 bar (0 ... 580 psi)

Output

 

Output signal

 
  • Two-wire system

4 ... 20 mA

  • 3-wire system

0 ... 20 mA

Measuring accuracy

to EN 60770-1

Linearity error including hysteresis (reference point adjustment)

= 0.2 % of full-scale value

Adjustment accuracy

= ± 0.2 % of full-scale value

Response time

< 20 ms

Influence of ambient temperature

 

On the enclosure

 
  • Zero

< 0.2 %/10 K of full-scale value

  • Measured span

< 0.2 %/10 K of full-scale value

On the process connection (remote seals)

Zero error (depends on design)

  • Flange remote seal
 
  • DN 25 / 1"

4.8 mbar/10 K (0.070 psi/10 K)

  • DN 32 / 1¼"

2.3 mbar/10 K (0.033 psi/10 K)

  • DN 40 / 1½"

1.6 mbar/10 K (0.023 psi/10 K)

  • DN 50 / 2"

0.6 mbar/10 K (0.009 psi/10 K)

  • Clamp-on seal
 
  • DN 25 / 1"

9.5 mbar/10 K (0.138 psi/10 K)

  • DN 32 / 1¼"

4.1 mbar/10 K (0.060 psi/10 K)

  • DN 40 / 1½"

3.9 mbar/10 K (0.057 psi/10 K)

  • DN 50 / 2"

3.9 mbar/10 K (0.057 psi/10 K)

The zero error specified for the process connection should be considered as a guideline for a standard design. We will produce a detailed system calculation on request. Systems with reduced remote seal errors are available on request.

Rated conditions

 

Installation conditions

 
  • Mounting position

Any, vertical as standard

Ambient conditions

 
  • Ambient temperature

-10 ... +70 °C (14 ... 158 °F)

  • Storage temperature

-10 ... +90 °C (14 ... 194 °F)

  • Process temperature

Max. 200 °C (392 °F), depending on design

  • Degree of protection (to EN 60529)

IP65, optional IP67

  • Electromagnetic compatibility
 
  • Emitted interference

To EN 50081 Part 1, issue 1993 (residential and industrial areas). The unit has no own emissions.

  • Noise immunity to

EN 50082 Part 2, issue March 1995 (industrial areas)

Construction

 

Weight (without remote seal)

 
  • Field housing

˜ 460 g (˜ 1.01 lb)

  • Enclosure with plug

˜ 200 g (˜ 0.44 lb)

Enclosure

 
  • Designs
  • Field housing IP65 or IP67, with screwed gland
  • Angled plug DIN 43650, IP65
  • Cable connection, IP67
  • Round plug connector M12, IP65
  • Material

Stainless steel, mat. No. 1.4404/316L/1.4305

Material of union nut

Polyamide (with electrical connection using plug or cable)

Electronics unit potted with silicone

Internal ventilation for measuring ranges < 16 bar (< 232 psi), through housing thread or connecting cable depending on design

Process connection

 
  • Versions

See ordering data

  • Material of coupling

Stainless steel, mat. No. 1.4404/316L

Power supply

 

Terminal voltage on transmitter

10 ... 30 V DC

Rated voltage

24 V DC

Certificates and approvals

 

Classification according to pressure equipment directive (DRGL 97/23/EC)

For gases of fluid group 1 and liquids of fluid group 1; complies with the requirements of Article 3, paragraph 1 (appendix 1); assigned to category III, conformity evaluation module H by the TÜV Nord

Explosion protection

 
  • Intrinsic safety "i"

TÜV 03 ATEX 2099 X

  • Identification

Ex II 2G EEx ib IIC T6

 

Share |
Video clips
Hot News
Certificates
Online Support
Sales Support
0901 185 205
Support via Skype     Support via Yahoo Messanger    
Technical Support
0938 77 40 99
Support via Skype     Support via Yahoo Messanger    
 
An Thinh Automation Ltd.,
26/50 Nguyen Minh Hoang street, Tan Binh District, HCM City, Viet Nam
Tel:(84) 08 3948 1715         Fax: (84) 08 3811 1332