Code:
7MF1562/7MF1564
Product name:
Z series for gage pressure
|
SITRANS P pressure transmitters, Z series for gage pressure
|
SITRANS P pressure transmitter, Z series for gage pressure |
Mode of operation |
|
Measuring principle |
Thin-film strain gauge |
Input |
|
Measured variable |
Gage pressure |
Measured range |
0 to 16 bar g (0 to 232 psi g) or
0 to 25 bar g (0 to 363 psi g) |
Output |
|
Current output signal |
4 ... 20 mA |
Measuring accuracy |
To EN 60770-1 |
Error in measurement (at 25 °C (77 °F), including conformity error, hysteresis and repeatability) |
0.5 % of full-scale value - typical |
Response time T99 |
< 0.1 s |
Long-term drift |
|
|
0.3 % of full-scale value/year - typical |
|
0.3 % of full-scale value/year - typical |
Influence of ambient temperature |
|
|
0.3 %/10 K (0.3 %/10 K) of full-scale value - typical |
|
0.3 %/10 K (0.3 %/10 K) of full-scale value - typical |
Rated conditions |
|
Medium conditions |
|
|
-30 ... +120 °C (-22 ... +248 °F) |
Degree of protection to EN 60529 |
IP65 |
Ambient conditions |
|
|
-25 ... 85 °C (-13 ... +185 °F) |
|
-50 ... 100 °C (-58 ... +212 °F) |
Design |
|
Weight |
≈ 0.2 kg (≈ 0.44 lb) |
Wetted parts materials |
|
|
Al2O3 - 96 % |
|
Brass, mat. No. 2.0402 |
|
Viton |
Process connection |
Male thread G½B female thread G1/8B |
Power supply |
|
Terminal voltage on pressure transmitter |
|
|
10 to 36 V DC |
Certificate and approvals |
|
Classification according to pressure equipment directive (DRGL 97/23/EC) |
For gases of fluid group 1 and liquids of fluid 1; complies with requirements of article 3, paragraph 3 (sound engineering practice) |
|
|
|