Mã sản phẩm:
DEPOLOX® 5 bare
Tên sản phẩm:
DEPOLOX® 5 bare electrode for Cl2, O3, ClO2 or KMnO4 measurement
|
The DEPOLOX® 5 measurement module can be used
with either the SFC electronic package for single
point analysis and control or the versatile MFC
electronic package for multiple measurements and
control.
|
General The DEPOLOX® 5 measurement module can be used with either the SFC electronic package for single point analysis and control or the versatile MFC electronic package for multiple measurements and control. It consists of a plug-in sensor card and flow cell with an integral, bare-electrode measurement configuration. Utilizing amperometric residual measurement technology, it is suitable for disinfection applications, ranging from simple measuring/ monitoring tasks to complex control processes for treating potable water, process water, and pool water. The DEPOLOX® 5 flow cell has a standard measuring range from 0.01 to 200 mg/l of free chlorine.
Typical applications
- Measurement and control tasks in potable water
- works and by potable water suppliers
- Process water monitoring in all water-based industrial processes
- Cooling water monitoring
Features The DEPOLOX® 5 flow cell consists of a 3-electrode system utilizing a bare-electrode design which provides a quick response time (90 % change < 20 sec.) with high accuracy (±2 % F.S.) when compared to membrane sensor technology. Hydrodynamic grit cleaning of the electrode surfaces maintains sensitivity to extend the intervals between calibrations. Zero point calibration is not necessary. A sample with a constant pH is required, however, the addition of a pH sensor can provide for compensated free chlorine measurement with the suitable electronic package. An integral multi-sensor provides a PT 1000 temperature measurement and monitors sample flow to provide a loss-of-flow alarm contact. The flow cell is supplied with a 1 m (3.3 ft.) screened coaxial cable. Up to three additional sensors can be fitted into the flow cell for measuring other water parameters. Utilizing “plug and play” technology allows the SFC or MFC controller to automatically recognize the sensor card and provide the correct display information. An analog output (0/4 to 20 mA) is available along with user configurable alarm contacts.
Benefits:
- Accurate measurement and high reproducibility
- Fast response time to meet fluctuating disinfection demands
- Rugged design and minimal maintenance due to hydrodynamic cell cleaning
- Economical, reagentless operation
- Proven operation in thousands of installations around the world
- Intuitive programming for user-friendly operation
|
|
|